Guralp-systems CMG-1T Bedienungsanleitung Seite 20

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 34
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 19
CMG-1T Ocean Bottom Seismometer Operator’s
Guide
Section 3, Page 4 September, 1999,
Issue C
The figures below gives the comparative response of a conventional velocity
output broadband sensor and hybrid output broadband sensor. The family of
curves shows the sensor output response to input acceleration in units of
V/m/s
2
. Curves 1, 2 and 3 are the conventional 360 sec, 100 sec and 30 sec
responses and curves 4 and 5 are the hybrid 10 sec and 30 sec responses.
Alternatively the system amplitude plots are given below as output against input
velocity.
Seitenansicht 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 33 34

Kommentare zu diesen Handbüchern

Keine Kommentare